FTV-1600C Series

Clamped Vortex Flowmeter
FTE-1600Q electromagnetic flowmeter adopts new type of sanitary lining material and lining technology, which meets the sanitary requirements of food industry. At the same time, it adopts stainless steel clamp connection to facilitate the rapid disassembly and cleaning of electromagnetic flowmeter, the electromagnetic flowmeter is not easy to be polluted in the use process, and can effectively prevent the accumulation of measuring fluid residues in the measuring tube, can be widely used in mineral water, soy sauce, jam, beer, fruit juice, rice wine, milk and other food and pharmaceutical manufacturing process and health, chemical and other fields.
Product Selection
Expert Q&A
5% off for new customers
Specification
Performance Features
The basic principle of the FTV-1600C series clamped vortex flowmeter is the Carmen vortex phenomenon. When the medium flows through a triangular cylinder-shaped vortex generator at a certain speed, two alternating rows of vortices are formed on its two sides, the vortex is called the Kármán Vortex Street. Strouhal built on the Carmen vortex theory by suggesting that the frequency of the Carmen vortex is proportional to the velocity of the fluid.
Corrosion-resistant stainless steel material, ABS full mold production probe, simple structure.
The detection sensor does not directly contact the measured medium, the performance is stable, and the life is long.
The output is a pulse signal proportional to the flow, without zero drift and with high precision.
Low pressure loss, low operating cost, more energy-saving significance.
The volumetric and mass flow rates of steam, gas and liquid can be measured.
The amplifier can be separated from the sensor (separation distance 15m) .
The use of noise elimination circuit and anti-vibration sensor head, so that the instrument has a certain anti-environmental vibration performance.
The pressure loss is small, the range of measurement is wide, the range is 10-40 times.
The temperature of the measurable medium reaches + 350 ° C 
No mechanical moving parts, long-term stability, simple structure for installation and maintenance.
The output of the sensor is a pulse frequency, its frequency is linear with the actual flow of the measured fluid, zero drift, performance is very stable.
The adoption of RS-485 communication, or Hart communication, can realize factory automation, integration.
Application:
FTV-1600C series clamp type vortex flowmeters are widely used to measure volume and mass flows of superheated steam, saturated steam, Compressed air and general gases and liquids.
Technical parameters
Measuring Range Liquid: 0.35 ~ 7.0 m/s gas: 5.0 ~ 60.0 m/s steam: 6.0 ~ 70.0 m/s
Suitable medium Ordinary gases, saturated vapors, superheated vapors, Compressed air, ordinary liquids
Nominal pipe diameter DN15-300MM (flange, clamp)
DN250-1800MM (insert)
Accuracy Level  ± 0.5% FS、 ± 1.0% FS、 ± 1.5% FS、 ± 2.5% FS
Repeatability error   ± 0.5% of the measured value
Range ratio 8:1    10:1    15:1
Display Widescreen LCD display, parameters and variables can be set through the panel keys (with password protection)
Instantaneous flow display, cumulative total display, percentage display, frequency display
M3/h、 kg/h、 t/h、 Hz
Operating Temperature -40-130 ° C (standard)
-40-350 ° C (high temperature type)
-40-80 ° C (explosion-proof)
Damping adjustment 0-50
Job Stress 1.6 mpa, 2.5 mpa, 4.0 mpa, special can be customized
Shell material Stainless steel, cast aluminum alloy
Straight pipe section requirements Entry: 15 * DN
Export: 5 * DN
Signal output  4-20mA, Pulse, RS485, Modbus protocol, Hart output
Power Supply  Pulse output: 12V DC, 24V DC (optional)
Two-wire system 4-20mA: 24V DC (with Hart Communication)
With 3.6 V lithium battery, service life of more than 2 years
Electrical Connection M20 * 1.5 terminal wiring
Connection mode  Standard flange, clamp, internal and external threads
Protection level   IP65(optional IP67)
Ambient temperature   - 30 degrees.-30 degrees. 60 degrees C...40 degrees C... 85 ° C (non-explosion-proof) , sensor-40 ° C... 80 ° C, converter-15 ° C... 60 degrees C
Relative humidity   5%~95%
Explosion-proof signs Explosion-proof ExdIICT5
Consume power <20W
Not sure which Instruments and sensors to choose?
No worries! We understand that getting the perfect Instruments and sensors that meets your needs can be pretty tricky. Our dedicated Engineering experts are here to support you with:
Expert advice on choosing the best material and structure to meet your needs
Professional guidance from engineers on building the best Sensors and Instruments experience
Anytime support with full-cycle project management and proactive updates
Get a product quote
Fill in our quote request form and we will reply to your message as soon as possible! Or send us an email at: info@doweston.com